Back to search
Published: 2017-05-10Captured: 2026-02-09Sysmex America IncProduct: Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan

FDA issued a Class II enforcement action. Affects Sysmex America Inc. Reason: Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protecti.

  • Classification: Class II
  • Firm: Sysmex America Inc
  • Status: Terminated
  • Distribution: Nationwide, Canada
Reason

Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of ...

Status:Terminated
Distribution:Nationwide, Canada
View Source26f311de